Power Supply |
100-120V or 220-240V, 50/60Hz or internal rechargeable battery with lifetime of 60 minutes |
|
Safety |
Class I, type CF |
|
LCD display |
Backlit high resolution 1024 x 600 10 inch touchscreen display 12 channel ECG waveform simultaneously |
|
Leads |
12 STANDARD or CABRERA leads |
|
Defibrillation protection |
Internal, according to current regulations |
|
Recording modes |
Manual, automatic sequential, automatic simultaneous, Rhythm lead selectable |
|
Print Channels |
12 channels , Rhythm lead selectable |
|
Writing system |
High-resolution thermal printer (8 dots/mm); Horizontal 40 dots/mm |
|
Thermal paper |
216mm wide 20M thermal paper in Z-fold packers or in rolls |
|
Paper speed |
5 - 6.25 - 10 - 12.5 - 25 - 50mm/s ± 3% |
|
Recording sensitivity |
Auto or Manual: 1.25 - 2.5 - 5 - 10 - 20 - 40mm/mV ±3% |
|
Input dynamics |
±500 mV @ 0 Hz ± 10 mV in the pass band |
|
Input impedance |
50 MΩ DC |
|
Power requirements |
AC220V/50Hz |
|
Common-mode rejection |
> 90 dB |
|
Frequency response |
0.05 to 150 Hz |
|
Time constant |
>3.3 s |
|
Data acquisition |
12 bits; 1000 samples/sec/channel when printing, filtering and computing |
|
Fuse Specification |
2Φ5x20mm |
|
Case storage |
150 cases at least, can be stored on external SD card |
|
Patient Cable |
standard 10 leads |
|
Resolution |
3.125 uV/bit |
|
Filters |
Low pass filter off 75 Hz, 100 Hz, 150 Hz |
|
Muscle filter: 25 or 35 Hz digital filters (-3 dB/oct) |
|
AC filter: digital notch filters 50 or 60 Hz selectable, - 20 dB or better |
|
Drift filter: 0.5 Hz high pass linear phase digital filter |
|
Net weight |
3.7kg (8.14 lb) |
|
Gross weight |
5.2kg (11.44lb) |