Standard with Fetal irritability, contractions and fetal heart, SM-110 Plus add blood oxygen, pulse rate and blood pressure. SM-110T Fetal Monitor for twins.
Description:
Unicare MCF-21K are high-tech fetal& maternal monitor, it offers a reliable, high-performance solution for fetal monitoring needs in doctors’ practices, clinics and hospitals, from pre-natal check-ups to antepartum monitoring of high-risk pregnancies
Feature:
-Standard configuration: FHR, TOCO,FM , AFM, FMP
-Folded 10.1 inch TFT LCD high-brightness screen
-12 elements water proof pulse doppler transducer
-Event marker for easy documentation of events and kick counts
-Fetal movement profile ( FMP) detects and automatically records gross fetal body movements
-Twin fetus monitoring capabilityand twin signal overlap alarm function
-Configuration audible and flashing alarms
-Built-in high speed thermal printer ( width: 112mm)
-Supporting TCP/IP or wireless connection to the central monitoring station
-Internal Li-ion battery power AC power from external power supply


Specifications:
Physical Specifications |
Dimension Weight |
Dimension |
32cm (W)*26cm (D)*9cm (H) |
Weight |
standard configuration: 2.8 kg |
Operation Environment |
Power requirements |
AC100-240V , 50/60Hz |
Power |
50VA |
Temperature |
5-40℃ |
Humidity |
30-93% non-condensing |
Fetal Performance Specifications |
Performance Specifications |
Display |
10.1inch (800X600 pixel) color TFT –LCD |
Indicator |
Visual on-display and integrated, alarm light, audible alarm |
Battery |
Internal rechargeable Li-ion battery power |
Alarm |
Visual on-display and integrated, alarm light, audible alarm |
Memory |
24-hours memory for fetal traces |
Printer |
America standard/international standard
Built-in thermal printer (112mm)
1/2/3cm/min real-time printing speed, fast print speed
(stored traces) |
Paper |
Z-fold, thermosensitive |
Record Message |
ID, name, age, date, time, print speed etc |
CMS |
Support connecting to central monitoring station through
LAN (TCP/IP) or wireless connection |
Fetal Technical Specifications |
Ultrasound (FHR) |
Measurement range |
US |
50 to 240 bpm |
Resolution |
Display |
1bpm |
Printer |
1bpm |
jitter@200 bpm |
≤3bpm |
Display update rate |
0.25second |
Beat to bea change(max.) for ultrasound |
28bpm |
US frequency |
1MHz±5%Hz |
TOCO |
Measurement method |
Strain gauge sensor element |
Sensitivity |
1 unit =2.5g |
Resolution |
1 unit |
Measurement range |
500 units |
Baseline setting |
0, 5,10, 15, 20 can be selective |
Update rate |
0.25 second |
Auto offset correction |
3 second after connecting the transducer |
Auto zero adjust
|
TOCO value is set to zero negative
following
a negative .Measure value for 5 seconds |
Auto
Fetal Movement (AFM) |
Technique |
Pulsed doppler ultrasound |
Range |
0-100% |
Resolution |
1% |
Automatic fetal movement actograph |
Manual fetal movement mark |
Maternal
ECG (MECG) |
Heart rate range |
30~300bpm |
Resolution |
1bpm |
Accuracy |
±1bpm/ ±1% |
Leads |
I, II, III, avL, avR, avF, V |
Sweep speed |
12.5mm/s, 25mm/s, 50mm/s |
Gain selection |
x 0.25, x0.5 ,x1, x2 |
Operation mode |
diagnostic, monitor, surgery |
Temperature (TEMP) |
Technique |
Thermistor probe CYSI 400 series compatible |
Measurement range |
0℃-50℃ |
Accuracy |
±0.1℃ |
|
|
|
|
|
|

|